PSA Testing methods, Siemens vs Abbott - Advanced Prostate...

Advanced Prostate Cancer

21,428 members26,837 posts

PSA Testing methods, Siemens vs Abbott

Birdwood profile image
3 Replies

Just recd latest results and noticed my previous test was done using the Roche Cobas method and all others with Siemens Centaur. Readings were 0.21 (S) 0.06 (R) and 0.19 (S) .... latest. Any possibility that the different methods are the reason for the tripling. Cheers

Written by
Birdwood profile image
Birdwood
To view profiles and participate in discussions please or .
Read more about...
3 Replies
Tall_Allen profile image
Tall_Allen

Yup.

Birdwood profile image
Birdwood in reply to Tall_Allen

Thanks. Was same pathology company, just different machines. Makes life difficult.

Justfor_ profile image
Justfor_

Definitely NO. I have used 6 different labs with whatever equipment they may have. Lab to lab, day to day a 20% discrepancy is not unusual, though not common. 30% calls for a retest by a third lab next day. Among the 50+ tests I have had up to now two errors of + 600% and - 300%.

You may also like...

Change in PSA testing method

Previous testing was centaur chemilumisometric immunoassay with PSA of 0.31. Fifteen weeks later,...

What is the minimum detection level of Ultrasensitive PSA test?

I just received my latest psa test result of 0.025 - 18 months after my SRT. The PSA seems to hang...

Post RP PSA vs uPSA test

three months and get a regular PSA test instead of the ultra-sensitive test because 0.2 will be the...

Confused by PSA tests -- Husband's PSA for past 3 yrs: <0.1; now (8/10/19) reported 0.05.

urologist, new testing lab, etc. It seems a different type of PSA test was just done with the...

PSA testing comparison question

blood, could different labs result in such a huge difference? I'm going to ask my Urologist office...